S E C T I O N 1 3
•
S p e c i f i c a t i o n s a n d A c c e s s o r i e s
38
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
SECTION 13
Specifications and Accessories
System Specifications
Standard Compliance
Guidelines
Product Safety
IEC 60601-1:2005
EMC
EN 60601-1-2, FCC 47 CFR Part 15,
Class B
Enclosure Protection
IEC 60529 IP Code: IPX1
Fluid Warmers
ASTM F2172-02
Physical
Dimensions
Height, Overall
24.1 cm
(9.5 inches)
Width, Overall
21.0 cm
(8.3 inches)
Depth, Overall
17.8 cm
(7.0 inches)
Weight, Dry
3.5 Kg
(7.6 lbs)
Weight, Wet (with recirculating
solution)
5.0 Kg
(11.0 lbs)
Weight, Shipping
3.6 Kg
(7.95 lbs)
Recirculating Solution
Capacity
1.4 L
(0.37 gallons)
Maximum Height on I.V. Pole
107 cm
(42 inches)
Environmental
Temperature
Humidity [%]
Operation
10
o
C to 45
o
C
10 to 95
Transportation
-18
o
C to 60
o
C
5 to 90
Storage
-18
o
C to 60
o
C
5 to 90
Thermal
Temperature
Temperature Set Point
40.0
o
C ± 0.1
o
C
Over Temperature Set Point
41.2
o
C
Electrical
Type
MAINS Power Input: 115V
115VAC, 50/60 H
Z
, 3.0 Amps
MAINS Auxiliary Supply Power
Output: 115V
115VAC, 50/60 H
Z