S E C T I O N 4
•
A s s e m b l y I n s t r u c t i o n s
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
11
Step 4 - Fill the Reservoir With Recirculating Solution
WARNINGS
WARNINGS
• Do not fill the HOTLINE
Warmer reservoir with a HOTLINE
Fluid Warming Set in place. Failure to remove the HOTLINE
Fluid Warming Set before the fill procedure may result in an air
lock in the HOTLINE
Warmer.
Recirculating Solution Protocols
Use one of the following solutions for the reservoir.
1
Prepare the recirculating solution.
2
Remove the fill-port plug (
a
).
3
Fill the reservoir with 1.4 liters of recirculating solution.
4
Replace the fill-port plug.
Step 5 - Perform the Electrical Safety Tests
Perform all applicable electrical safety tests as required per institutional
procedure. Refer to Section 9,
Testing
, for more information about
electrical safety testing.
Recirculating Solution
Preparation
Maintenance
0.3% Hydrogen Peroxide
Solution
Mix 140 ml of 3% hydrogen
peroxide with 1,260 ml of
distilled water.
Replace solution and
disinfect reservoir every
12 months.
Distilled Water
Use distilled water.
Replace solution and
disinfect reservoir every
30 days.
35% Isopropyl Alcohol
Solution
Mix 700 ml of 70% isopropyl
alcohol with 700 ml of distilled
water.
Replace solution and
disinfect reservoir every
30 days.
Note: Use distilled water only, not tap water. Failure to do so may cause build-up of
mineral deposits in the recirculating solution path, which may impair heater
performance.
1.4 LITERS
a