S E C T I O N 8
•
T r o u b l e s h o o t i n g
22
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
Hot cabinet
Check for blocked air vents on the bottom or the back of
the HOTLINE
Warmer.
Note: Room temperature above 40
°
C may cause the
HOTLINE
Warmer to shut down and the Over
Temperature alarm to activate. In this situation, turn
the power switch OFF and allow the HOTLINE
Warmer to cool down before returning it to service.
Difficult to install the
HOTLINE
Fluid
Warming Set
Lubricate O-rings in the socket. Refer to Section 10,
Maintenance
, for the procedure.
Recirculating
solution leaks at the
socket where the
HOTLINE
Fluid
Warming Set plugs
into the HOTLINE
Warmer
Replace O-rings. Use the O-ring Replacement Kit: P/N
80-04-001. Refer to Section 10,
Maintenance
, for the
procedure.
Electrical
interference -
receiving or
transmitting
1 Move the HOTLINE
Warmer away from the device in
question.
2 Plug the HOTLINE
Warmer into a separate electrical
circuit.
• If the problem continues, notify Smiths Medical or
your local Smiths Medical distributor.
Problem
Check the following