S E C T I O N 6
•
O p e r a t i o n
14
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
Power and Alarm Test Panel
The Power and Alarm Test Panel is located on the left side of the
HOTLINE
Warmer next to the reservoir. This panel contains two
pressure-sensitive buttons that are activated when pressed, and the
ON/OFF switch.
1
Over Temperature Alarm Test Button - The Over Temperature
Alarm Test is used to confirm the proper operation of the Over
Temperature circuitry.
2
Alarm Test Button - The Alarm Signal Test is used to confirm
proper operation of the visual and audible alarms.
3
Power ON/OFF Switch - The black switch toggles to turn power
ON and OFF.
Reservoir Level Display
The reservoir for the recirculating solution is located on the left side of
the HOTLINE
Warmer, next to the Power and Alarm Test Panel. The
level of the recirculating solution is visible in the reservoir. Two symbols
indicate the maximum (
a
) and minimum (
b
) solution level
requirements.
1.4 LITERS
a
b