S E C T I O N 7
•
O p e r a t i n g I n s t r u c t i o n s
20
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
Continued from previous page
WARNINGS
WARNINGS
• Not for use with pressure devices generating over 300 mmHg.
Pressure greater than 300 mmHg may compromise the integrity
of the HOTLINE
®
Fluid Warming Set.
1
Wait until the recirculating solution temperature display reaches
39
°
C, which indicates the HOTLINE
Warmer is ready for use.
2
Adjust the rate of I.V. flow using the clamp on the intravenous
administration set.
Note: Do not kink the
Disposable Set. Do not restrict the circulation
of the solution through the tubing.
Step 5 - After Use
WARNINGS
WARNINGS
• Blood and blood products could contain pathogenic organisms.
Failure to follow institutional policy and procedures for
biomedical-hazardous materials could lead to exposure to
harmful pathogens.
1
Turn OFF the power switch.
2
Remove the
HOTLINE
Fluid Warming Set, and insert the reflux
plug (
a
) into the socket.
3
After use, handle and dispose of the
HOTLINE
Fluid Warming
Set in a safe manner according to local guidelines for disposal of
contaminated medical waste.
4
Wipe down the external surfaces of the HOTLINE
Warmer with
mild liquid detergent soap and warm tap water mixture and a soft
cloth or sponge. See Section 10,
Maintenance,
for more details
about cleaning and external disinfection.
Storage
Store the HOTLINE
Warmer in a cool, dry place. Do not expose to
extreme temperatures. See Section 13,
Specifications
, for more details.
o
o
C
C
HOTLINE
Fluid Warmer
a