Rev.
00
GW2045 – GW1060 Manual
Pag. 25 di 43
KEY TO PROGRAMMES AND DEFAULT DOSE TABLE
SYMBOL
DESCRIPTION
P1, P2, P3
Peristaltic pumps
DD
:
Detergent measuring device
N.C.
Heating not controlled
COMPONENT
STANDARD DOSE
P1/DD
5ml/litre
P2.
4ml/litre
P3
9ml/litre
WATER INLET
GW2045
GW1060
STANDARD PHASE
5 litres
9 litres
WASH AND NEUTRALISATION PHASE
4 litres
6 litres
DEMINERALISED WATER RINSE
5 litres
9 litres
NOTE
If the demineralised water attachment has not been selected, the phase is
replaced by one using cold water.
Approximate times
programmes 1 to 8: About 10 mins, programmes 2, 3, 4, 5, 6, 7: About 1 hour
The given times are approximate and related to the three-phase 400V version.
PERSONALISABLE PROGRAMMES 6 AND 7
Programmes 6 and 7,
CUSTOM 1
and
CUSTOM 2
respectively, can be personalised
by authorised SMEG technicians, on the customer’s request.
By default, programme 6 corresponds to
“thermal disinfection at 93°C for 3’
”.
By default, programme 7 corresponds to
“thermal disinfection at 93°C for 10’ with a
prewash”.
Keep track of the specifications for any personalised programmes, using the
blank table given below.