SETUP MENU
SETUP
menu, available functions:
1.
Position 1
: For
dose of detergent
in pump P1. Set the ml/litre required as dose of detergent with
maximum threshold allowed at 20ml/litre. In 1ml steps (increase/decrease by using buttons (5) and
(4)). P1 Default dosage is 5 ml/litre.
2.
Position 2:
Similar to Position 1 for
P2
, neutraliser. P2 Default dosage is 3 ml/litre.
3.
Position 3:
Similar to Position 1 for
P3
. P3 Default dosage is 0 ml/litre.
4.
Position 4:
Setting of
hardness of water
in °f, from 5 to 60 with 5°f steps. Default is a w ater
hardness of 40°f. Important parameter for the corre ct functioning of softener incorporated into
appliance.
5.
Position 5
: Choice for
automatic opening of door
at end of cycle. “ON” as default. ON and OFF
appear on the display in segments to select the two options.
6.
Position 6
:
Printer language
- to be selected from the various languages available; default is
English.
Selected Language
German
French
Spanish
Italian
English
Abbreviation on Display
dE
Fr
ES
It
En
7.
Position 7:
Demineralised water
attachment, yes/no.
Demineralised water attachment
Present
Absent
Abbreviation on Display
dn
--
8.
Position 8:
Date-time
(for details se paragraph DATE AND TIME SETUP).
It is possible to access and change the following parameters using programme selection and the
buttons (5) and (4):
Setup field
Year
month
day
Time
Minutes
Seconds
Inactive
position
Inactive
position
Switch
position
1
2
3
4
5
6
7
8