4.4
Interchangeability
In order to obtain an accurate and better temperature compensated response, each sensor is submitted
to a characterization process and the specific data is stored in an EEPROM located in the sensor body.
The main board, in this operation, reads the sensor serial number and compares it with the number
stored in the main board. In case they do not match, the circuit considers that the sensor has been
changed and will probe the memory of the new sensor for the following information:
✓
Temperature compensation coefficients.
✓
Sensor trim data, including 5-point characterization curve.
✓
Sensor characteristics: type, range, diaphragm material and fill fluid.
Information not transferred during sensor replacement will remain unchanged in the main board
memory. Thus, information such as Upper Value, Lower Value, Damping, Pressure Unit and
replaceable transmitter parts (Flange, O-ring, etc.) shall be updated, depending whether the correct
information is that of the sensor or the main board. In the case of a new sensor, the main board will
have the most updated information; in the opposite case, the sensor will have the correct information.
Depending on the situation, the updating shall be from one or the other.
Returning Materials
Should it become necessary to return the transmitter and/or configurator to
SMAR
, simply contact our
office, informing the defective instrument serial number, and return it to our factory.
If it becomes necessary to return the transmitter and/or configurator to Smar, simply contact our office,
informing the defective instrument's serial number, and return it to our factory. In order to speed up
analysis and solution of the problem, the defective item should be returned with the Service Request
Form (SRF
– Appendix B) properly filled with a description of the failure observed and with as much
details as possible. Other information concerning to the instrument operation, such as service and
process conditions, is also helpful.
Instruments returned or to be revised outside the guarantee term should be accompanied by a purchase
order or a quote request.
Summary of Contents for LD290
Page 1: ......
Page 2: ......
Page 6: ...LD290 Operation and Maintenance Instruction Manual VI...
Page 9: ...Installation 1 3 Figure 1 1 b Dimensional Drawing and Mounting Position for LD290 Sanitary...
Page 11: ...Installation 1 5 Figure 1 1 d Dimensional Drawing and Mounting Position for LD290 Level...
Page 18: ...LD290 Operation and Maintenance Instruction Manual 1 12...
Page 32: ...LD290 Operation and Maintenance Instruction Manual 3 10...
Page 42: ...LD290 Operation and Maintenance Instruction Manual 4 10...
Page 61: ...Appendix A A 7 CEPEL Centro de Pesquisa de Energia El trica...
Page 62: ...LD290 Certifications Information A 8...
Page 63: ...Appendix A A 9 FM Approvals...
Page 64: ...LD290 Certifications Information A 10...
Page 66: ...LD290 Appendix B B 2...