SLE2000 HFO Service Manual
Doc No. SM/0001: Issue 7.0
Page 71 of 84
7. O
PERATIONAL
V
ERIFICATION
7.1 B
ASIC
O
PERATION
With mains power connected, switch unit on and observe that:
1)
All audible alarms sound for approximately 2 seconds.
2)
All warning LED’s except ‘BLOCK’ and ‘HFO FAIL’ are activated for approximately 2 seconds
3)
All 7 Segment displays increment from 0 to 9 (Check all segments are operative)
4)
LCD firmware initialises & displays SLE logo.
5)
With mode switch set successively in CPAP, CMV, PTV, and SIMV, disconnect the mains supply
& confirm that the Mains Failure audible alarm is activated in each position. On completion re-
apply mains power.
6)
By shorting together both pins of XL1 on the A0702/02 PCB, check operation of the processor
watchdog timer by observing a processor reset function together with audible and visual alarms.
7)
By shorting together both pins of XTL1 on the JO 736 PCB, check operation of the processor
watchdog timer by noting audible & visual alarms. Select CPAP mode, HFO OFF.
8)
Check operation of MANUAL BREATH push-button, and by adjustment of INSP TIME control
confirm that a range of inspiratory times from 0.1 sec. to 3.0 sec. shall be obtained, and that this
adjustment can be monitored on the INSP TIME display.
9)
Check that no mandatory breaths are present other than those produced by the MANUAL
BREATH push-button.
10)
Switch HFO successively to EXP, INSP, and CONT.modes. Ensure that manual breaths are not
delivered in any HFO modes but OFF.
11)
Set pressure display switch to MIN position. By selecting a range of inspiratory pressures from 0
cmH
2
O to 60 cm H
2
O in 10 cmH
2
O steps, ensure accuracy of digitally displayed pressure to less
than 1 cmH
2
O error against an RT 200 reference.
12)
By adjusting HIGH alarm setting to displayed inspiratory pressure, and operating MANUAL
BREATH push-button, observe that inspiratory plateau on LCD pressure display and bargraph
display correspond with alarm line and LED Segment, respectively.
13)
Check that HIGH audible and visual alarm operates at pressures above or at HIGH alarm setting,
and fresh gas solenoid is de-energised at pressures greater than 2cm above HIGH alarm setting.
Summary of Contents for SLE2000 HFO
Page 27: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 27 of 84 ...
Page 29: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 29 of 84 5 2 A0701 02 ...
Page 30: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 30 of 84 ...
Page 32: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 32 of 84 5 3 A0702 02 ...
Page 33: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 33 of 84 ...
Page 36: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 36 of 84 5 4 A0736 01 ...
Page 37: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 37 of 84 ...
Page 40: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 40 of 84 5 5 A0737 01 ...
Page 41: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 41 of 84 ...
Page 43: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 43 of 84 5 6 A0738 01 ...
Page 44: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 44 of 84 ...
Page 46: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 46 of 84 5 7 A0738 02 ...
Page 47: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 47 of 84 ...
Page 49: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 49 of 84 5 8 A0739 01 ...
Page 50: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 50 of 84 ...
Page 52: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 52 of 84 5 9 SERIAL INTERFACE OPTION ...
Page 54: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 54 of 84 5 10 WIRELOOM ...
Page 55: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 55 of 84 5 11 FRONT PANEL ...
Page 58: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 58 of 84 5 14 ELECTRONIC CHASSIS ...
Page 59: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 59 of 84 ...