SLE2000 HFO Service Manual
Doc No. SM/0001: Issue 7.0
Page 67 of 84
Turn motor speed control fully anti-clockwise
Verify that frequency at TP1 is: .......... 600Hz±60Hz(1.67ms±0.167ms)Factory set.
Connect a patient circuit and test lung to the ventilator.
Adjust HFO peak to peak pressure to: ........................................20cmH
2
O approx.
Monitor buffered transducer output using a calibrated oscilloscope.
Verify that minimum HFO rate is:................................................3Hz ,333ms±3ms
and that maximum HFO rate is: ................................................ 20Hz ,50ms±0.5ms
6.2.5 HFO M
AIN
B
OARD
(A0736/01)
Connect Pressure source to PTR1
6.2.5.1 Scaling of pressure LCD display
Set up a pressure of:.......................................................... 60 cmH
2
O ±0.5cmH
2
O.
Adjust P14 to give:........................................................................... 4.63V±0.01V
Set up a pressure of:.......................................................... -60 cmH
2
O ±0.5cmH
2
O
Confirm TP2 is:................................................................................ 0.28V±0.01V
Remove pressure source from PTR1 port.
Set low alarm to: ..................................................................................... 0 cmH
2
O
Confirm that the displayed low alarm level coincides with the displayed pressure level.
Reconnect pressure source to PTR1 port.
Set high, FTC and low alarm controls fully clockwise. Switch ventilator to CMV mode.
Adjust P11 so that displayed high alarm level:........................................ 62 cmH
2
O
Adjust P13 so that displayed FTC level:................................................. 50 cmH
2
O
Adjust P12 so that displayed low alarm level: ........................................ 40 cmH
2
O
Confirm that the fresh gas dump solenoid is de-activated at a position above the high
alarm setting of between:................................................................... 0cm and 2cm
6.2.5.2 Bargraph Calibration
Set high pressure limit to : ..................................................................... 60cmH
2
O
Set FTC pressure limit to : ..................................................................... 50cmH
2
O
Summary of Contents for SLE2000 HFO
Page 27: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 27 of 84 ...
Page 29: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 29 of 84 5 2 A0701 02 ...
Page 30: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 30 of 84 ...
Page 32: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 32 of 84 5 3 A0702 02 ...
Page 33: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 33 of 84 ...
Page 36: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 36 of 84 5 4 A0736 01 ...
Page 37: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 37 of 84 ...
Page 40: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 40 of 84 5 5 A0737 01 ...
Page 41: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 41 of 84 ...
Page 43: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 43 of 84 5 6 A0738 01 ...
Page 44: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 44 of 84 ...
Page 46: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 46 of 84 5 7 A0738 02 ...
Page 47: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 47 of 84 ...
Page 49: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 49 of 84 5 8 A0739 01 ...
Page 50: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 50 of 84 ...
Page 52: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 52 of 84 5 9 SERIAL INTERFACE OPTION ...
Page 54: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 54 of 84 5 10 WIRELOOM ...
Page 55: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 55 of 84 5 11 FRONT PANEL ...
Page 58: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 58 of 84 5 14 ELECTRONIC CHASSIS ...
Page 59: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 59 of 84 ...