SLE2000 HFO Service Manual
Doc No. SM/0001: Issue 7.0
Page 65 of 84
6.2.3
V
ENTILATOR
CPU B
OARD
(A0702/02)
6.2.3.1 U18/26(ADC)
Monitor U18/26 (ADC)
With PTR1 ports open to atmosphere adjust RV2 to give: ............................ 3.11V
Set up a pressure of: .............................................................................. 60cmH
2
O
Adjust RV5 so that U18/26 is: ..................................................................... 4.91V
Set up a pressure of: ............................................................................. -60cmH
2
O
Check that U18 /26 is: ................................................................................. 1.28V
6.2.3.2 Buffered transducer output.
Monitor PLB/23b:
With PTR1 ports open to atmosphere adjust RV3 to give: ............................ 3.42V
Set up a pressure of:............................................................................... 60cmH
2
O
Adjust RV4 to give: ..................................................................................... 6.42V
Set up a pressure of:.............................................................................. -60cmH
2
O
Check that the buffered transducer output (PLB/23b) is:................... 0.42V±0.01V
6.2.3.3 Fresh Gas Block and Leak Alarm
Connect DVM between pins 2 and 4 of PTR2
With PTR2 ports open to atmosphere adjust RV500 to give: .............. 0.00V±1mV
Connect the fresh gas pressure tube to +ve port of PTR2
Switch ventilator on to CPAP mode and confirm that fresh gas flow is :5L/min ±
0.25L/min
With a suitable flow control valve create a pressure at the fresh gas outlet of:
........................................................................................ 17.5cmH2O±0.5cmH2O
Adjust RV501 so that the leak alarm and LED just trip on.
By adjustment of the flow control valve create a pressure at the fresh gas outlet of:
.......................................................................................... 80cmH
2
O ± 0.5cmH
2
O
Adjust RV502 so that the high pressure alarm and LED just trip on.
6.2.3.4 Maximum Inspiration Time
Select CPAP mode
With max. insp. time control turned fully clockwise adjust RV6 to display :3.00 secs.
Summary of Contents for SLE2000 HFO
Page 27: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 27 of 84 ...
Page 29: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 29 of 84 5 2 A0701 02 ...
Page 30: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 30 of 84 ...
Page 32: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 32 of 84 5 3 A0702 02 ...
Page 33: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 33 of 84 ...
Page 36: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 36 of 84 5 4 A0736 01 ...
Page 37: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 37 of 84 ...
Page 40: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 40 of 84 5 5 A0737 01 ...
Page 41: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 41 of 84 ...
Page 43: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 43 of 84 5 6 A0738 01 ...
Page 44: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 44 of 84 ...
Page 46: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 46 of 84 5 7 A0738 02 ...
Page 47: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 47 of 84 ...
Page 49: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 49 of 84 5 8 A0739 01 ...
Page 50: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 50 of 84 ...
Page 52: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 52 of 84 5 9 SERIAL INTERFACE OPTION ...
Page 54: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 54 of 84 5 10 WIRELOOM ...
Page 55: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 55 of 84 5 11 FRONT PANEL ...
Page 58: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 58 of 84 5 14 ELECTRONIC CHASSIS ...
Page 59: ...SLE2000 HFO Service Manual Doc No SM 0001 Issue 7 0 Page 59 of 84 ...