43
Unlocking the Power Drive System
Turn the unit ON and notice the sweep
action of the LEDS. Press the HORN
BUTTON two (2) times in ten (10)
seconds. The LEDS will flash and be
steady ON. The drive may be operated.
SW Scale Meter (optional)
The SW Scale Meter is an optional feature with a one touch
patient weigh system. It is only available for the Bari Rehab
Platform™.
NOTE:
It is recommended to have the bed in the lowest
position and casters locked in any direction when weighing.
Weighing the patient with the SW Scale Meter:
1.
Turn on Scale and Display Weight
Press “On Weight”.
The meter turns off automatically after one minute.
The weight reading is not saved.
2.
Zero Scale
Press and hold “On Weight” and “Zero”.
NOTE:
Never “ZERO” the scale with the patient in the bed.
“RECALIBRATING ZERO SCALE: - 00 –” appears in the Display Window.
Release buttons and the scale is zeroed.
NOTE:
The scale cannot be zeroed while in the kg mode.
3.
Before Transport – Hold Weight Reading
Press “On Weight” to weigh the patient and display the weight.