Sirona Dental Systems GmbH
12 Assistant element
Operating Instructions C4
+
12.1 Assistant element
59 06 313 D 3382
D 3382.201.01.13.02
65
12
Assistant element
12.1
Assistant element
The assistant element can be swiveled to the side and to
the front.
To do this grasp the assistant element laterally at the
handpiece rack (1).
The ease of the swivel movement towards the front can
be set with the knob (2).
CAUTION
Make sure that the backrest does not collide with the
assistant element if you have positioned the latter in the
movement range of the backrest.
Also make sure that the assistant element is not posi-
tioned above the patient when the chair moves upward
(on contact with the patient, a safety switch stops the
movement).
Removing a handpiece rack with closed suction hand-
piece holders is described in the document
“Care and
Cleaning by the Practice Team”
.
DANGER
The suction hose handpieces contain magnets. In
the case of body contact of the suction hose hand-
piece in the vicinity of a cardiac pacemaker, this can
have a disturbing effect on the functioning of the car-
diac pacemaker.
Therefore avoid bringing the suction hose hand-
piece into the vicinity of a cardiac pacemaker.
Furthermore, direct contact of the suction hose
handpiece with magnetic cards can lead to data loss.
Suction hose handpieces with integrated magnets
are marked with the following symbol: