11 Basic treatment center settings in the Mode dialog
Sirona Dental Systems GmbH
11.19 Mode: WHITESCREEN YES – NO
Operating Instructions C4
+
59 06 313 D 3382
64
D 3382.201.01.13.02
11.19
Mode: WHITESCREEN YES – NO
Presetting a white screen for the SIVISION
monitor
If
YES
flashes –
The
“X-ray image viewer”
key switches ON/OFF both
the X-ray image viewer and a
WHITE SCREEN
on the
SIVISION 3 monitor.
If
NO
flashes –
The
“X-ray image viewer”
key only switches the X-ray
image viewer ON and OFF.
If an X-ray image viewer and SIVISION 3 are installed,
always select
“NO”.
11.20
Mode: A/B VIDEO MODE ON – OFF
Locking the video mode switchover
This mode is only displayed in connection with an inte-
grated SIVISION 3.
If
ON
is flashing –
You can use the
“A/B”
key to switch between VIDEO
mode and PC mode after picking up the SIVISION 3
camera.
If
OFF
is flashing –
You cannot use the
“A/B”
key to switch between VIDEO
mode and PC mode for SIVISION 3.
This prevents the selected mode from being changed
inadvertently.
11.21
Mode: BLOW OUT SPRAY
Prevents drop formation
You can activate the instrument blow-out function in this
setting option. This removes any residual water (one to
two drops) by blowing air through the instrument heads.
Just press
the
–
or
+
key to switch this function
ON
or
OFF.
The selected item flashes.
NOTICE
i
This function is switched
OFF
in the factory setting.
NOTICE
i
Sirona recommends activating this function especially
for the C3
+
treatment center.
A
WHITESCREEN
YES – NO
WHITESCREEN
YES – NO
A
A / B VIDEO MODE
ON – OFF
A / B VIDEO MODE
ON – OFF
A/B
A
BLOW OUT SPRAY
YES – NO
BLOW OUT SPRAY
YES – NO