11 Basic treatment center settings in the Mode dialog
Sirona Dental Systems GmbH
11.13 Mode: # FUNCTION TOGGLE - MOMENTARY
Operating Instructions C4
+
59 06 313 D 3382
62
D 3382.201.01.13.02
Setting the torque of the electric motors
This Mode dialog is used to set the automatic controller
action of the SL or SL ISO electric motors. Individual set-
tings are possible for each motor holder.
First a message prompting removal of an instrument
appears.
After removing the instrument, select the setting dialog
by pressing the Mode key again:
MAX TORQUE
If
Yes
is flashing:
The motor will attain a higher torque at speeds > 15,000
rpm.
If
No
is flashing:
The running smoothness of the motor will be higher at
speeds > 15,000 rpm.
NOTICE
i
The new setting becomes effective only after the corre-
sponding motor is removed again.
NOTICE
i
11.13
Mode: # FUNCTION TOGGLE
−
MOMENTARY
Free function key # as button or switch
In this mode you can preset whether you want to use the
free function key
#
as a button or as a switch (see page
17).
11.14
Mode: PURGE TIME... SEC
Setting the purge time for flushing the water
paths
(Only for dentist element)
You can preset a purging time between 60 and 120 sec-
onds for each instrument with the
–
/
+
keys for the
purge function
(see page 54).
A
TAKE INSTRUMENT
A
MAX. TORQUE
YES – NO
MAX. TORQUE
YES – NO
A
# FUNCTION
TOGGLE – MOMENTARY
# FUNCTION
TOGGLE – MOMENTARY
A
PURGE TIME
120 SEC
PURGE TIME
120 SEC