Sirona Dental Systems GmbH
9 Dentist element and instrument functions
Operating Instructions C4
+
9.10 Satelec Mini L.E.D. curing light
59 06 313 D 3382
D 3382.201.01.13.02
49
•
1100mW/cm² (± 10%) with the standard light guide,
dia. 7.5mm.
•
2000mW/cm² (± 10%) with the optional booster light
guide, dia. 5.5mm.
4.3 Pulsed curing mode
This mode is selected by pressing the function button.
The light operates at full power in “pulsed” mode; it emits
radiation in 10 consecutive light pulses of 1s each.
Between the individual pulses there is a 250 ms break.
LED 4.3 shows your selection.
4.4 “Soft start” mode
This mode is selected by pressing the function button.
The “soft start” mode provides:
•
A 10-second “soft start” from 0 to 1100 mW/cm², or
from 0 to 2000 mW/cm² with the (optional) “booster
light guide”, dia. 5.5 mm.
•
Full power during 10 seconds.
LED 4.4 shows your selection.
Handling
After you switch on the treatment center, the
Mini L.E.D.
is by default in the quick curing mode (4.2).
If you change the operating mode of the
Mini L.E.D.
while the treatment center is in operation, the new mode
remains valid even after placing the
Mini L.E.D.
in its
holder.
After you have selected a mode, the
Mini L.E.D.
is ready
for operation.
Position the light guide as close as possible to the com-
posite material surface you want to photopolymerize.
Make sure that the light guide never touches the material
to be cured, as this may cause damage to the light guide
and reduce its effectiveness.
•
Never work without the glare shield (5).
•
Never look into the light which is reflected by the
tooth surface.
0