11.12 Instrument settings via Mode dialog: REMOVE INSTRUMENT.
59 06 313 D 3382
D 3382.201.01.13.02
61
If you want to reset the camera to its factory settings,
select
HOME
with the
– / +
keys. When you confirm by
pressing the Mode key,
HOME
is displayed in bold type.
To exit this Mode dialog, select
EXIT
and press the
Mode key, or put the camera back in place. Automatic
exit from the dialog is disabled as soon as
USER
or
HOME
is selected. This allows you to repeat the white
balance as often as necessary until you obtain the
desired result.
11.12
Instrument settings via Mode dialog: REMOVE INSTRUMENT.
This Mode dialog enables the following settings
• Instrument light voltage
and
• Maximum torque of SL motors
Setting the instrument light voltage
You can change the voltage setting for instruments in
this Mode dialog.
First a message prompting removal of an instrument
appears.
Then the current voltage setting of the (removed) instru-
ment is displayed.
Using the
-/+
keys, you can adjust the instrument light
voltage from 3.0 to 5.6V in steps of 0.1V.
Starting at 3.9V,
WARNING
also appears.
After depositing the instrument, please remove and set
the next instrument.
Observe the voltage specified by the instrument
manufacturer!
NOTICE
i
The voltage (3.6V) is already preselected for Sirona in-
struments.
A
TAKE INSTRUMENT
A
MAX LIGHT VOLT
3.9 V WARNING
MAX LIGHT VOLT
3.9 V WARNING