10 Operation light, tumbler filling, cuspidor flushing, purge
Sirona Dental Systems GmbH
10.5 Purge / Auto Purge function (purging the water paths)
Operating Instructions C4
+
59 06 313 D 3382
54
D 3382.201.01.13.02
10.5
Purge
/
Auto Purge function (purging the water paths)
PURGE function
(only for dentist element)
The purge function enables you to flush the water paths.
To do this, press the quick setting key 1 (> 3 s) when all
instruments are in their holders.
You will be prompted to choose between PURGE
/
APURGE
and SANITATION
1
.
Select PURGE with the
–
or
+
key. PURGE must flash.
If you press quick setting key 1 briefly (< 3 s), the follow-
ing prompt will appear if all instruments are in their hold-
ers:
Pick up the instruments you want to flush and hold them
above the cuspidor.
The maximum water flow must be set on the instru-
ments.
Press quick setting key 1 again
(< 3 s), the instruments
are then purged one after the other.
The purging time in seconds set previously in the Mode
dialog (11.14) runs down for each instrument individu-
ally.
An audible signal sounds after the last instrument has
been flushed.
Finally place the instruments back in their holders.
1
A
SELECT MODE
PURGE / APURGE / SAN
SELECT MODE
PURGE / APURGE / SAN
1. The SANITATION function is described in the instructions
“Care and Cleaning by the Practice Team”.
A
SELECT ITEMS
FOR PURGING
SELECT ITEMS
FOR PURGING
max.
1