32 Settings on Dentist and Assistant Elements
Sirona Dental Systems GmbH
59 58 470 D 3370
44
D 3370.031.01.10.02
Braking effect at the head of the dentist element
At the customer's request, the braking effect of the rotary
joint at the head of the dentist element can be adjusted.
4.
Remove the covering of the support arm.
NOTE
i
To illustrate the position of the setscrews
A
more clearly, the
brake is shown without attachment parts in figure X.
5.
Adjust the setscrews
A
equally until the braking force
corresponds to the customer's wishes.
6.
Reattach the covering.
Figure X
A
A
A
Torx 15
Torx 15
Torx 15
4./6.
5.
5.
5.
5.