22 Attaching the Water Unit Housing Covers
Sirona Dental Systems GmbH
59 58 470 D 3370
28
D 3370.031.01.10.02
1.
Remove the lower part of the amalgam separator or of
the automatic separator (if installed).
2.
Loosen screws
A
. Slide both aluminum profiles approx.
10 mm outwards.
3.
Place the housing cover at an angle on the foot from
above and push down.
4.
Slide both aluminum profiles over protrusions
B
in
housing cover. Fasten screws
A
tight.
5.
Insert the two side covers in tongue-and-groove joint.
6.
Remove the cover from the front housing.
Attach front housing cover, inserting it in side covers.
First screw on upper screw
D
loosely (lining up coupling
nut if necessary).
ATTENTION
Make sure that cables and hoses are not pinched behind
the domed nuts for screws
E
.
7.
Screw in two lower screws
E
.
Fasten all screws tightly.
22 Attaching the Water Unit Housing Covers
2.
A
A
D
E
B
1.
2.
3.
4.
4.
7.
6.
5.