Sirona Dental Systems GmbH
28 Compare and adjust operating air pressures
59 58 470 D 3370
D 3370.031.01.10.02
37
båÖäáëÜ
NOTE
i
The operating pressures for the instruments are set before
they leave the factory. The settings for propellant air and
spray air can be changed if the customer so desires.
Precise adjustment of the propellant and spray air pres-
sures is possible using the 4-on-1 probe (Order No. 36 66
336).
Factory settings for running turbine:
Turbines:
T1 Control, TC1 and TC3, TM1 TITAN:
Propellant air (shown on the display)
3.5 bar
±
0.1 bar
Propellant air at 4-on-1 probe (TL)
2.7 bar
±
0.1 bar
Spray air at 4-on-1 probe (SL)
2.7 bar
±
0.1 bar
EL 1, motor SL
The cooling air pressure for the motor results from the driv-
ing air setting of the turbine. A turbine has to be fitted
instead of the motor for setting. For setting refer to page 38.
NOTE
i
TL = propellant air
SW = spray water
SL = spray air
RL = return air
28 Compare and adjust operating air pressures
TL
SW
SL
RL