scorpION3 5200969 Rev 4
27
Table 2. Typical Setup Recommendations in 0.5 m/s (90 fpm) Airflow
Table 3. Typical Setup Recommendations: in still air (no airflow)
[Not recommended for use at less than 300 mm (12 inches) ]
Measures of performance should be made using the test methods
described in either your company methods and standard practice or
industry standard ANSI/ESD STM 3.1-2006 Ionization, Section 6.2
laminar flow hood ionization. A Charge Plate Monitor that meets the
requirements of ESD STM 3.1 should be used for this purpose.
Approx.
Distance
Offset
Volts
Discharge
Time (sec)
Bar
Mode
Pulse
Time (sec)
Pulse
Overlap %
Output
Level %
300 mm
(12”)
15
6
SS
-
-
30
450 mm
(18”)
10
10
SS
-
-
40
30
9
pulse
0.50
0
40
60
7
pulse
1.00
0
40
600 mm
(24”)
10
15
SS
-
-
50
30
12
pulse
0.75
30
50
60
10
pulse
1.25
30
50
750 mm
(30”)
10
18
SS
-
-
60
30
15
pulse
1.00
50
60
60
12
pulse
1.50
50
60
Approx.
Distance
Offset
Volts
Discharge
Time (sec)
Bar
Mode
Pulse
Time (sec)
Pulse
Overlap %
Output
Level %
300 mm
(12”)
±
50
30
pulse
0.50
0
30
±
100
15
pulse
1.00 0
30
450 mm
(18”)
±
50
30
pulse
1.50
30
40
±
100
20
pulse
1.75 30
40
600 mm
(24”)
±
50
45
pulse
2.50
50
50
±
100
30
pulse
3.00
50
50
750 mm
(30”)
±
50
60
pulse
4.00
70
60
±
100
40
pulse
5.00
60
60
Summary of Contents for scorpION3
Page 1: ...I O N I Z A T I O N S O L U T I O N S Ionizing Bar System scorpION3TM User s Manual...
Page 6: ...scorpION3 5200969 Rev 4 1 1 Description 1 1 scorpION3 Ionizing Bar System 1 2 Features...
Page 29: ...scorpION3 5200969 Rev 4 24...
Page 30: ...scorpION3 5200969 Rev 4 25 3 Operation 3 1 Settings...
Page 41: ...scorpION3 5200969 Rev 4 36...
Page 42: ...scorpION3 5200969 Rev 4 37 5 Specifications 5 1 Specifications 5 2 Parts Accessories...
Page 49: ...scorpION3 5200969 Rev 4 44...