To add a value in the table of calibration points, use the
keys, enter the calibration points
(real value and value read) then confirm using the VAL key.
•
Repeat this operation for all the calibration points (maximum of 4).
To delete a line, select it then use the
key.
To edit a line, select it then use the navigation key (
) to make editing possible.
•
Confirm using the
VAL
key to return to the measurement screen.
To ensure the measurements are made using the calibration coefficients defined earlier, go to the
configuration/function
menu.
•
In the
thermocouple
field, select sensor1 (SONDE1-SENSOR1 below).
Note: the Calibrated sensors are at the top of the list and their name is preceded by a *.
•
Confirm the latter using the
VAL
key.
The chosen calibrated sensor is displayed in the measurement screen.
Name of
calibrated sensor
used