Sensor
Range of measurement Resolution Accuracy
K
-240 up to -50 °C
-50 up to 0 °C
0 up to 1372 °C
0.2 °C
0.1 °C
0.05 °C
0.60 °C
0.10 °C
0.013 % of rdg. + 0.08 °C
T
-240 up to -100 °C
-100 up to 0 °C
0 up to 400 °C
0.2 °C
0.05 °C
0.05 °C
0.40 °C
0.10 °C
0.013 % of rdg. + 0.08 °C
J
-210 up to 0 °C
0 up to 1200 °C
0.05 °C
0.05 °C
0.20 °C
0.013 % of rdg. + 0.07 °C
E
-240 up to -100 °C
-100 up to 40 °C
40 up to 1000 °C
0.1 °C
0.1 °C
0.05 °C
0.25 °C
0.10 °C
0.013 % of rdg. + 0.05 °C
R
-50 up to 350 °C
350 up to 900 °C
900 up to 1768 °C
0.5 °C
0.2 °C
0.1 °C
0.5 °C
0.013 % of rdg. + 0.35 °C
0.013 % of rdg. + 0.2 °C
S
-50 up to 120 °C
120 up to 450 °C
450 up to 1768 °C
0.5 °C
0.2 °C
0.1 °C
0.8 °C
0.013 % of rdg. + 0.35 °C
0.013 % of rdg. + 0.25 °C
B
400 up to 850 °C
850 up to 1820 °C
0.2 °C
0.1 °C
0.013 % of rdg. + 0.4 °C
0.013 % of rdg.+ 0.2 °C
U
-200 up to 600 °C
0.05 °C
0.15 °C
L
-200 up to 900 °C
0.05 °C
0.20 °C
C
-20 up to 900 °C
900 up to 2310 °C
0.1 °C
0.1 °C
0.25 °C
0.013 % of rdg.+ 0.15 °C
N
-240 up to -190 °C
-190 up to -110 °C
-110 up to 0 °C
0 up to 1300 °C
0.2 °C
0.1 °C
0.05 °C
0.05 °C
0.3 °C
0.15 °C
0.08 °C
0.013 % of rdg. + 0.06 °C
Platinum
-100 up to + 1 400 °C
0.05 °C
0.3 °C
Mo
0 up to + 1375 °C
0.05 °C
0.013 % of rdg. + 0.06 °C
NiMo/NiCo -50 up to + 1410 °C
0.05 °C
0.013 % of rdg. + 0.30 °C
The accuracy is guaranteed for a cold junction (JR) at 0 °C.
With use of the internal JR (except thermocouple B) add an additional uncertainty of 0.3 °C
at 0 °C.
For other temperatures, temperature thermocouple sensitivity must be taken into account,
therefore uncertainty at Temperature T is 0.3 °C*S(0 °C)/S(T).
•
Temperature coefficient: < 10 % of the accuracy / °C.
•
Display in °C, °F and K.
•
It is possible, thermocouple B excepted, to choose the location of the cold junction by
programming from the keyboard:
- external at 0 °C,
- internal (compensation for the temperature of the terminals of the unit),
- by programming the temperature.