2.3.6
Frequency generation (signal)
•
Display the
TRANSMISSION CONFIGURATION
dialogue box.
•
Select the
Frequency
transmission function, then the
range using the function and navigation keys.
•
Select the
Signal
"Output type".
•
Enter the amplitude of the signal between 0 and 20 V.
•
Confirm with ENTER.
The following ranges are available:
Range
1000 Hz
10 000 Hz
Resolution
0,1 Hz *
1 Hz *
Max. Amplitude
20 V
20 V
* Caution:
The detected value may differ from the displayed value. The generated frequency is represented by
a fixed frequency whose value is divided by an integer. Therefore, the displayed value (with the res-
olution being near the display) is this newly calculated value which most gets closer to the detect-
ed value.
The display units may be in Hz or Beats per minute (BPM).
Connection of the frequency source generated is between the Hz and COM terminals.
2.3.7
Frequency generation (dry contact)
•
Display the
TRANSMISSION CONFIGURATION
dialogue box.
•
Select the
Frequency
transmission function, then the
range using the function and navigation keys.
•
Select the "Type of Output"
Dry contact
.
•
Confirm with ENTER.
The following ranges are available:
Range
1000 Hz
10 000 Hz
Resolution
0,1 Hz *
1 Hz *
Max. Amplitude
20 V
20 V
* Caution:
The detected value may differ from the displayed value. The generated frequency is represented by
a fixed frequency whose value is divided by an integer. Therefore, the displayed value (with the res-
olution being near the display) is this newly calculated value which most gets closer to the detect-
ed value.
The display units may be in Hz or Beats per minute (BPM).
Connection of the frequency source generated is between the Hz and COM terminals.