EC RTD.2
Technical data
Technical changes reserved
- 33 -
Characteristics
EC RTD.2
Input
RTD (Cu10)
-
Input range
-
Resolution
-
Accuracy*
-
Remarks
-10.0…250.0
Ω
0.1 C
± (0.05 % of r 0.6 °C)
Note 7 & 8
RTD (Cu50)
-
Input range
-
Resolution
-
Accuracy*
-
Remarks
-50.0…150.0
Ω
0.1
Ω
± (0.05 % of r 0.6 °C)
Note 7 & 8
RTD (Pt10 (385))
-
Input range
-
Resolution
-
Accuracy*
-
Remarks
-200.0…850.0
Ω
0.1
Ω
± (0.05 % of r 0.6 °C)
Note 7 & 8
RTD (Pt1000 (385))
-
Input range
-
Resolution
-
Accuracy*
-
Remarks
-200.0…850.0
Ω
0.1
Ω
± (0.05 % of r 0.6 °C)
Note 7 & 8
RTD (Pt200 (385))
-
Input range
-
Resolution
-
Accuracy*
-
Remarks
-200…630
Ω
0.1
Ω
± (0.05 % of r 0.6 °C)
Note 9
RTD (Pt500 (385))
-
Input range
-
Resolution
-
Accuracy*
-
Remarks
-200…630
Ω
0.1
Ω
± (0.05 % of r 0.6 °C)
Note 9
RTD (Pt1000 (385))
-
Input range
-
Resolution
-
Accuracy*
-
Remarks
-200…630
Ω
0.1
Ω
± (0.05 % of r 0.6 °C)
Note 9
Note 7: Exciting current is set as: ±0.5…±3 mA
Note 8: If the exciting current is set as ±0.1…0.5 mA, add an extra 0.5 °C to additional error.
Note 9: Exciting current is set as± 0.05…±0.3 mA