EC RTD.2
Technical data
Technical changes reserved
- 31 -
9
Technical data
The technical data of customised versions may differ from the data in these instructions.
Please observe the information specified on the type plate.
Accuracy specified at 23 °C ±5 °C & 75 % RH for a period of one year after calibration.
9.1
Characteristics EC RTD.2
Characteristics
EC RTD.2
Output
Resistance (400
Ω
)
-
Output range
-
Resolution
-
Accuracy*
-
Remarks
0.0…400.0
Ω
0.1
Ω
± (0.05 % of r 0.2
Ω
)
Note 1 & 2
Resistance (4000
Ω
)
-
Output range
-
Resolution
-
Accuracy*
-
Remarks
0…4000
Ω
1
Ω
± (0.05 % of r 2
Ω
)
Note 1
RTD (Cu10)
-
Output range
-
Resolution
-
Accuracy*
-
Remarks
-10…250 °C
0.1 °C
± (0.05 % of r 0.6 °C)
Note 1 & 3
Employs Pt(385)
RTD (Cu50)
-
Output range
-
Resolution
-
Accuracy*
-
Remarks
-50.0…150.0 °C
0.1 °C
± (0.05 % of r 0.6 °C)
Note 1 & 3
Employs Pt(385)
RTD (Pt10 (385))
-
Output range
-
Resolution
-
Accuracy*
-
Remarks
-200.0…850.0 °C
0.1 °C
± (0.05 % of r 0.6 °C)
Note 1 & 3
Employs Pt(385)
* The accuracy does not include lead resistance.
Note 1: Exciting current is set as: ± 0.5…±3 mA
Note 2: If the exciting current is set as ±0.1…0.5 mA, add additional 0.1
Ω
to error.
Note 3: If the exciting current is set as ±0.1…0.5 mA, add additional 0.5 °C to error.