Output from EC RTD.2
EC RTD.2
- 26 -
© SIKA
•
Ba_EC-RTD.2
•
12/2018
Power on:
NOTICE!
To ensure proper operation, please wait 5 seconds before turning the instrument on af-
ter changing the batteries.
Press the power button to turn on the device.
When the device is turned on, it starts to make self-diagnosis internally and the full screen is
in display. After this, appropriate operation should be carried out.
Power off:
Press the power button for one second to turn off the device.
NOTICE!
To ensure correct operation of the device, it is recommended to turn off the device for
5 seconds and then restart the instrument.
Automatic Power-off:
The device will automatically turn itself off after 15 minutes without any action. Follow the
steps below to disable the Auto Power-off function:
1. Turn off the device.
2. Press the power button to turn on the device.
3. When the LCD shows all its contents, press the RANG button
The device enters into the setting mode, and the symbol AP-XX appears on the LCD.
4. Press the
▼
button to activate or deactivate the power-off function.
AP-OF indicates that automatic power-off is disabled.
AP-ON indicates that automatic power-off is enabled.
5. Press the 4/3/2W button to save the setting.
6. Turn off the device again to exit from the setting function.
5
Output from EC RTD.2
The output terminal of the instrument can generate resistance or thermal resistance ana-
logue signals set by the user.
CAUTION! Material damage!
If any improper voltage is applied to the output terminal, it will cause damage to the in-
ternal circuit.
Do not any apply voltage to the output terminal during the operation.