*3-9900.090*
Signet 9900 Transmitter
3-9900.090 Rev. G 02/17
English
The 9900 Transmitter, a member of Signet's line of SmartPro
®
instruments,
provides a single-channel interface for all Flow, pH/ORP, Conductivity/Resistivity,
Salinity, Pressure, Temperature, Level, Dissolved Oxygen, Turbidity, Batch and
other applications.
The 9900 is available in either Panel or Field Mount. Both versions run on
10.8 to 35.2 VDC power (24 VDC nominal), and can power certain sensors on
loop power (see NOTE on page 11).
The 9900 Transmitter, also allows third-party 4 to 20 mA signals to be used as an
input (optional Signet 8058 i-Go
®
Signal Converter required, sold separately).
Your new Signet 9900 Transmitter needs to be calibrated and the
sensor needs to be initialized prior to use. The following steps
outline the recommended procedure to start up a new system.
1. Module Installation (page 3)
2. Installation (page 7)
3. Wiring (page 8)
4. Sensor Wiring (page 11)
5. Power Wiring (page 17)
6. Relay and Open Collector Wiring (page 18)
7. Relay Functions (page 19)
8. Operation (page 23)
9. Menu System (page 25)
Look for the
Quick Start
icon to quickly
set up your
new 9900.
Panel Mount
Operating Instructions
Field Mount
The 9900 is compatible with all GF Signet
products listed in the column to the right.
●
pH and ORP electrodes require the
Signet 2750/2751 DryLoc
®
Sensor
Electronics (sold separately).
●
Conductivity/Resistivity or Salinity
measurement requires either the optional
Direct Conductivity/Resistivity Module
(part number 3-9900.394) or the Signet
2850 Conductivity/Resistivity Sensor
Electronics (sold separately).
NOTE:
If using the 2850, use the one-
channel Digital (S
3
L) models. The two-
channel model 3-2850-63 may be used
with only one channel connected. Do
not use with both channels connected.
The 4 to 20 mA models 3-2850-52 and
3-2850-62 are incompatible with the 9900.
●
Turbidity measurement using Signet 4150
or Dissolved Oxygen measurement using
Signet 2610-31 requires Signet 8058 i-Go
Signal Converter (sold separately).
Flow
515*/8510*, 525*, 2000,
2100, 2507, 2536*/8512*,
2537, 2540*, 2551, 2552
pH/ORP
2724-2726 with 2750*/2751
2734-2736 with 2750*/2751
2756-WTx–2757-WTx with
3719 and 2750*/2751
2764-2767 with 2750*/2751
2774-2777 with 2750*/2751
Conductivity/Resistivity, Salinity
2819-2823 with
2850 or Cond/Res Module
2839-2842 with
2850 or Cond/Res Module
Level, Temperature, Pressure
2250*, 2350*, 2450*
Turbidity
4150 requires 8058
Dissolved Oxygen
2610-41 direct to 9900
2610-31 requires 8058
* Can be run on Loop Power
(see NOTE on page 11)