57
9900 Transmitter
Manual Calibration Procedure - pH
Requires prepared buffers. System calibration is possible with two known pH
solutions within 0 to 14 pH (buffers of pH 4.01, 7, or 10 are recommended,
but use a buffer close to your own process value.)
Single-point calibration sets
STANDARD only; Signet
recommends a two-point
calibration to set SLOPE in
addition to STANDARD.
Quick Manual
Calibration Procedures:
1-Point Calibration
:
1. Set solution standard.
2-Point Calibration
(recommended):
1. Set solution standard.
2. Set solution slope.
ENTER
30s
30s
ENTER
ENTER
ENTER
+
Allow for stabilization
30 seconds to
several minutes
to accept
Display returns to
VIEW Mode
To Calibrate:
To Change Reading:
To Set Slope:
To Set Calibration Date:
Place electrode tip in
pH buffer.
Place electrode tip in
pH buffer two pH units
different than standard.
to accept
to accept
to accept
Allow for stabilization
30 seconds to
several minutes
To Change Reading:
Calibration Procedures - pH