42
9900 Transmitter
TEMPERATURE Setup Checklist
1. Make sure TEMPERATURE sensor type is selected
(see System Setup Menu, page 25).
2. If LOOP is used, set the minimum and maximum 4 to 20 mA setpoints.
3. Set Units of Measurement (°C or °F).
4. Set Last Cal Date and initials.
5. If desired, set up relay functions for your own application.
This is the normal display
and does not time out.
CAL Menu
ON prevents relays from activating while making adjustments, and relays in PULSE mode will suspend
pulsing. Output is held until the user exits the CAL menu. Select OFF/ON.
Default = OFF.
Provides a maximum 20 °C offset to match to a known standard (external reference).
Resets Temperature Calibration to factory settings. After pressing
►
, select YES/NO.
Enter date of calibration (mm-dd-yyyy) and initials of calibrator (ii).
INPUT Menu
Enter string up to 13 characters (optional).
Default = “TEMPERATURE”.
Select
°
C or
°
F.
Default =
°
C.
Dampens display, output and relay response rates.
Select Low, Med, High, OFF. (See discussion in Appendix.)
Default = OFF.
VIEW Mode Menu
Displays the 4 to 20 mA LOOP output
Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays.
Displays remaining time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE:
(Reset the Timer) In the RELAY menu, select TEST RELAY function.
The timer will reset to 0 if the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.
Temperature