MA MCTS II, Ver. 1.4
11
11/2018
SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: [email protected], Web: www.signaltec.com
7.4. Burden module specifications
The transducer system delivers the transducer output current at
the 4 mm output terminals on the back panel of the rack. For those
instruments which
don’t have current input terminals, optional high
precision passive and active plug-on burden modules with very low
phase angle error are available. The active voltage output modules
are supplied by the MCTS rack with a 3-pole D-SUB connector.
7.4.1. Passive plug-on burden resistor specifications
Passive plug-
on burden resistors are available from 1 Ω
to 50 Ω. The burden resistor is limited by the transducer
and the length of the connection cable. For higher output
voltages active plug-on burden amplifiers are available.
Order
Number
Resistance
Value
Accuracy
Max. Output
Voltage
Bandwidth
Phase
Error
Load
Influence
MCTS/BR1/0.02
1 Ω
0.02 % of MV
1.00 V
rms
@ 1000 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
MCTS/BR1.5/0.02
1.5
Ω
0.02 % of MV
1.00 V
rms
@ 667 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
MCTS/BR2.5/0.02
2.5 Ω
0.02 % of MV
1.58 V
rms
@ 632 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
MCTS/BR3.75/0.02
3.75 Ω
0.02 % of MV
1.94 V
rms
@ 516 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
MCTS/BR5/0.02
5 Ω
0.02 % of MV
2.24 V
rms
@ 447 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
MCTS/BR10/0.01
10 Ω
0.01 % of MV
3.16 V
rms
@ 316 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
MCTS/BR25/0.01
25 Ω
0.01 % of MV
5.00 V
rms
@ 200 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
MCTS/BR50/0.01
50 Ω
0.01 % of MV
7.07 V
rms
@ 141 mA
rms
> 1 MHz
< 1⁰ @ 100 kHz
< 0.1 ppm/mW
Mechanical Data
Width:
Height:
Depth:
Mass:
51 mm
51 mm
62 mm (Connectors included)
85 g