MA MCTS II, Ver. 1.4
8
11/2018
SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: [email protected], Web: www.signaltec.com
7.
Specifications
7.1 . Rack specifications
Front
Rear
Dimensions
Cabinet Width:
Cabinet Height:
Cabinet Depth:
442 mm (19” rack type)
87,2 mm (2 HU)
415 mm
Mass Basic Unit:
Mass Internal power supply:
Mass 19” mounting brackets:
6.8 kg
0.5 kg
0.5 kg
General Data
Operation Temperature:
Operation Humidity:
Warranty period:
-
10 … 60 ⁰C
20 … 90 % RH, noncondensing
36 Months
To ensure mechanical stability please use 19” rack rails.
Electrical Data
Supply voltage:
Output voltages:
Max. Power Consumption:
100-240 VAC, 50/60 Hz, fused with 6 A slow blow
+ 15 V
DC
, 3.4 A
max
/ -15 V
DC
, 2.0 A
max
The maximum power consumption depends on the number of channels
installed, the number of transducers connected and the current consumption
of each transducer. The maximum power consumption of a six channel MCTS
with transducers IN 2000-S at 2000 A
rms
primary current is around 175 W.