6 Using the centrifuge
40
Version 08/2012, Rev. 1.6 of 04/02/2014 • sb
Translation of the original operating manual
Pos: 150 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/Spinco ntr ol S ( Zent 7)/ 060 -00 30- 0020 -0 050 Me nü Sta nda rd -nu r KS @ 1 0\m od_1 328 697 106 319 _68. docx @ 58 068 @ 4 @ 1
6.3.2.5 Standard menu
The standard menu is displayed a few seconds after the centrifuge has
been switched on. The following parameters can be displayed and
modified.
Abb. 17: Standard menu
Speed
In the upper section of the field, the set speed of the centrifuge is displayed.
The actual speed is displayed below this value. The values are stated in
revolutions per minute (rpm) and depend on the RCF values (see chapter
2.2.2.1 - "
Speed, radius, and relative centrifugal force
"). The maximum
speed values depend on the rotor that is used.
Relative centrifugal force (RCF)
The relative centrifugal force is the acceleration that the sample is
subjected to. The set value of this parameter is displayed in the upper
section of this field, with the actual value shown below. The values are
stated in g (gravitational acceleration) and depend on the speed value (see
chapter 2.2.2.1 - "
Speed, radius, and relative centrifugal force
"). The
maximum RCF values depend on the rotor that is used.
Temperature
The set temperature is displayed in the upper section of the field, with the
current rotor chamber temperature shown below. Temperatures between
-20 °C and +40 °C can be preselected.
NOTE
The centrifuge is not equipped with an active heater. As a result,
temperatures above room temperature depend on the air friction of the
running motor.