5 Set-up and connection
Version 08/2012, Rev. 1.6 of 04/02/2014 • sb
27
Translation of the original operating manual
Pos: 93 / 100 Sigma /100 BA Ze ntrifu gen (Sta nda rdm odul e)/0 50 Auf stellun g u nd Ansc hluss/ 050 A ufstellu ng und An schluss == == === == == === == == === == == === == == == @ 7\ mod_ 130 630 929 901 0_6 8.doc x @ 4 345 7 @ 1 @ 1
5 Set-up and connection
Pos: 94 / 100 Sigma /110 BA Ze ntrifu gen (Proj ekte )/8KS_8KBS/05 0 Aufst ellun g un d Anschl uss/0 50- 001 0 Aufst ellor t @ 7 \mo d_1 307 4331 569 87_ 68. docx @ 442 19 @ 2 @ 1
5.1 Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted
to ambient air.
• Air cooled centrifuges: Ensure sufficient ventilation.
NOTE
The centrifuge 8KS / SKBS with an air-cooled compressor should not be
set up with its left side against a wall, since otherwise the hot air, which is
emitted out the back, will be drawn in again as fresh air for cooling. As a
result, the unit will switch off due to overheating (see chapter 7.2 - "
Table
of error codes
").
• Water cooled centrifuges: Provide sufficient water throughput.
• Do not position the centrifuge near heat generators.
• Avoid direct sunlight (UV radiation).
• During transport from cold to warmer places, condensational water will
collect inside the centrifuge. Allow sufficient time for drying (min. 24 h)
before using the centrifuge again.
Pos: 95 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 96 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 97 / 100 Sigma /110 BA Ze ntrifu gen (Proj ekte )/8KS_8KBS/05 0 Aufst ellun g un d Anschl uss/0 50- 001 1 Aufst ellen und Ausrich ten der Ze ntrif uge @ 1 3\mo d_1 341 390 884 312 _68. docx @ 68 411 @ 2 @ 1
5.2
Set-up and alignment of the centrifuge
DANGER
Ensure that the centrifuge is disconnected from the power supply during
the set-up and alignment.
NOTE
ay attention to the layout plan (see chapter 11.7 - "
Layout plan
")!
Pos: 98 / 010 Unive rsalm odul e/ Seit enwec hsel @ 0\m od_ 120 211 6244 312 _0. docx @ 10 5 @ @ 1