corresponding RCF value. The deceleration (brake) and acceleration
curves correspond to curve 9 and the runtime field indicates “infinite”
(continuous run).
Fig. 15: "RAPID_TEMP" program
NOTE
The program will only be loaded if the actual temperature is below the set
temperature.
• Press the start button in order to start the temperature control run.
• During the temperature control run, only the set temperature can be
changed.
The current status of the program will be displayed on the status bar.
The “RAPID_TEMP“ program will be stopped under the following
conditions:
The set value is reached. In this case, the “RAPID_TEMP” program
stops with a sound signal, and the standstill heating system will be
activated.
• The stop button is pressed. The “RAPID_TEMP” program will be
stopped prematurely. No message will be issued when the set
temperature is reached.
• A parameter is changed (except for the temperature) or another input is
made. In this case, the “RAPID_TEMP” program will be aborted. No
message will be issued when the set temperature is reached.
After the stop, the previous program will be reloaded or the changed
parameters will be adopted as the new settings.
NOTE
The automatic lid opening function is suppressed after a temperature
control run in order to prevent the material from cooling again.