6-16HS
6 Using the centrifuge
28
Version 06/2012, Rev. 1.0 • sb-kue
Translation of the original operating manual
Pos: 110 /10 0 Sigm a/11 0 BA Z entrif uge n (Pr ojekt e)/6 -16S _6- 16HS/06 0 Bet rieb/ 060 -00 20- 002 0-0 030 6-1 6S Einsetz en vo n Zube hör @ 1 3\m od_ 134 0359 480 750 _68. docx @ 67 730 @ 4 4 @ 1
6.2.2.3
Installation of accessories
• Only use inserts that are suitable for the rotor (see chapter 11.1 -
"
Range of accessories
").
• All buckets of the swing-out rotor need to be installed when spinning.
• Always load the opposite inserts/buckets of the rotors with the same
accessories and fill to avoid imbalance.
Centrifugation with different tube sizes
Working with different tube sizes is possible. In this case, however, it is
very important that opposite inserts are identical (see figure).
Fig. 7: Permissible loading of the swing-out rotor with different tube sizes
Centrifugation with low capacity:
The tubes must be installed symmetrically so that the buckets and their
inserts are loaded evenly (see figure).
Fig. 8: Permissible loading of a swing-out rotor
NOTE
The buckets 13621 and 13420 must be installed in sets of 4 and in the
prescribed order (see the numbers on the buckets and on the rotor). The
buckets are dynamically balanced at the factory. Mixed use with only two
buckets on opposite sides will cause increased imbalance.
• When installing the 500 ml bottles use the supplied supporting rings