Version 08/2008, Rev. 1.7 - sb
page 42 of 75
English translation
)
Info!
The automatic lid opening function is suppressed after a rapid cooling phase in
order to prevent the system from reheating.
)
Attention!
The temperature value indicates the air temperature in the rotor chamber. Please
bear in mind that the rotor – depending on its type – will reach this temperature
only after a corresponding precooling time.
•
Open the lid in order to insert the samples. The refrigeration unit will stop so
that no white frost can form inside the rotor chamber.
5.3.5 Rotors
The available rotors for the centrifuge can be selected here (rotor 12124 in the
example). This is only possible at a standstill.
To select a rotor:
•
Turn the left rotary knob until “set” appears in front of the option “Rotor” on
the display (Fig. 5.20).
•
Select the option by pressing or turning the right rotary knob. "Set" will now
flash.
•
Turn the right rotary knob until the desired rotor number is displayed.
•
Press the right rotary knob or the start key to confirm the input. If this is not
done, the value will be automatically reset to the last setting.
Fig. 5.20:
Preselection of a rotor