Version 08/2008, Rev. 1.7 - sb
page 23 of 75
English translation
4
Set-up and Connection
4.1 Unpacking the Centrifuge
The centrifuge is packaged in a slip-lid box.
•
Remove the upper part of the box.
•
Take out the box containing the accessories.
•
Remove the upper foam cushion.
•
Remove the slip-lid box.
•
Lift the centrifuge upwards with a lifting device or with a sufficient number of
people to lift it safely. When lifting or carrying the centrifuge, always reach
under the centrifuge from the side.
)
Attention!
The centrifuge weighs approx. 41 kg!
Please retain the packaging for any possible future transport of the centrifuge.
4.1.1
Transport Safety Device
The transport safety device of the SIGMA 1-15PK consists of a screw which is
located at the bottom panel.
)
Attention!
The transport safety device must be removed prior to start-up because the
screw locks the motor bearings!
Procedure:
•
Lift the centrifuge upwards at the front side. Always reach under from the
side.
•
Put a suitable object, e.g. a wooden block, between table and centrifuge. A
coloured plastic screw can now be seen at the bottom panel.
•
Unscrew the screw by hand anti-clockwise.
•
Please retain the transport safety device for the possibility of the return of the
centrifuge.