![background image](http://html1.mh-extra.com/html/sigma/1-14/1-14_operating-manual_1267872025.webp)
1-14K
Using the Centrifuge
Version 05/2010, Rev. 1.8 of 01/09/2011- sb
25
Translation of the original operating manual
6.2.2.3
Installation of Accessories
•
Only use inserts that are suitable for the rotor (see 11.1 "Suitable
Accessories").
•
Always load the opposite inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
Centrifugation with low capacity: The tubes must be installed
symmetrically so that the buckets and their inserts are loaded evenly.
fig. 6.2: Permissible loading of an angle rotor
6.2.2.4
Tubes
•
Load the tubes outside of the centrifuge. Liquids in the bores of
the rotor cause corrosion .
•
Fill the tubes carefully and arrange them according to their
weight. Imbalances result in the excessive wear of the bearings.
•
Observe the instructions on safety and hazards in chapter 3!