![background image](http://html1.mh-extra.com/html/sigma/1-14/1-14_operating-manual_1267872017.webp)
1-14K
Safety
Version 05/2010, Rev. 1.8 of 01/09/2011- sb
17
Translation of the original operating manual
3.6.5
Safety Instructions for Centrifugation
For safe operation, observe the following before starting the
centrifuge:
•
Ensure that the centrifuge was set up properly (see chapter 5
"Set-up and Connection“).
•
Maintain a safety distance of at least 30 cm (12 inches) around
the centrifuge.
•
Do not store any dangerous goods in the safety area of the
centrifuge.
•
Do not stay in the safety area longer than what is absolutely
necessary for the operation of the centrifuge.
•
Only use the centrifuge with rotors and accessories that have
been approved by the manufacturer. We explicitly warn against
the use of equipment of poor quality. Breaking glass or bursting
vessels can cause dangerous imbalances at high speeds.
•
Ensure that the rotor is correctly fitted (see 6.2.2.1"Installation of
the Rotor“).
•
Observe the instructions on the installation of accessories (see
6.2.2.3).
•
The rotor must be loaded symmetrically at equal weights.
•
If liquids with a density
>
1.2 g/cm
3
are used, reduce the speed
(see 11.3.2 "Density").
•
Do not use the centrifuge if the rotor is loaded asymmetrically.
•
Do not use the centrifuge with tubes that are excessively long.
3.6.6
Resistance of Plastics
Chemical influences have a strong effect on the polymeric chains of
plastics, and therefore, on their physical properties. Plastic parts can
be damaged if solvents, acids, or alkaline solutions are used.
•
Refer to the resistance data (see chapter 11)!
3.6.7
Service Life of Rotors and Accessories
The rotors and accessories have a limited service life.
•
Perform regular checks (at least once per month) for safety
reasons!
•
Pay special attention to changes, such as corrosion, cracks,
material abrasion etc.
•
Aluminium rotors must be scrapped after 50,000 cycles or
max. 7 years.
•
If other data concerning the service life are engraved on the rotor
Summary of Contents for 1-14