1-14K
Maintenance and Service
38
Version 05/2010, Rev. 1.8 of 01/09/2011- sb
Translation of the original operating manual
Check the material regularly (at least once a month) for
- cracks
- visible damage of the surface
- pressure marks
- signs of corrosion
- other changes.
•
Check the bores of the rotors.
•
Replace any damaged components immediately for your own
safety.
•
Grease rotor, lid seal and adapters at least once a week with
slushing oil (part no. 70104).
8.1.4
Microhematocrite Rotor
•
Remove the microhematrocrite rotor for cleaning.
•
Wipe the centrifuge chamber clean.
•
Replace the rubber ring in the event of wear or glass breakage
(part no. 16002 for rotor 11026).
8.2 Sterilization and Disinfection of the Rotor Chamber and
Accessories
•
Use commercially-available disinfectants such as, for example,
Sagrotan
, Buraton
, or Terralin
(available at chemist’s shops or
drugstores).
•
The centrifuges and the accessories consist of various materials.
A possible incompatibility must be considered.
•
Before using cleaning or decontamination agents that were not
recommended by us, contact the manufacturer to ensure that
such a procedure will not damage the centrifuge.
•
For autoclaving, consider the continuous heat resistance of the
individual materials (see 8.2.1 “Autoclaving“).
Please contact us if you have any queries. (see 7.3 "Service Contact").
If dangerous materials (e.g. infectious and pathogenic substances)
are used, the centrifuge and accessories must be disinfected.