7.2 Error Codes
Error no. Kind of error
Measure
Note
1-9
System error
•
allow to slow down
•
power off/on
All these errors stop the
centrifuge or cause it to
decelerate brakeless
10-19
Speedometer error
•
allow to slow down
•
power off/on
20-29
Motor error
•
power off/on
•
ensure ventilation
30-39
EEPROM error
•
allow to slow down
•
power off/on
With error 34, 35, 36, the
centrifuge will stop;
with error 37, 38 error
message only
40-45
Temperature error
(only for refrigerated
centrifuges)
•
allow to slow down
•
power off/on
•
allow to cool down
•
provide better ventilation
46-49
Imbalance error
•
allow to slow down
•
power off
•
eliminate the imbalance
50-59
Lid error
•
press lid key
•
close lid
•
remove foreign matter from the
opening of the lid lock device
With error 50 and 51, the
centrifuge will stop
60-69
Process error
•
allow to slow down
•
power off/on
With error 60 message
"power failure during run",
with error 61, message "stop
after power on"
70-79
Communication error
•
allow to slow down
•
power off/on
80-89
Parameter error
•
allow to slow down
•
power off/on
•
provide for better ventilation
With error 83, error
message only
90-99
Other errors
•
check connections
Fig. 7.4: Error codes from software version 009
If it is impossible to eliminate the errors, contact the service!