2.11.2 TRIGger:DELay {<seconds>|MIN|MAX|DEF}
TRIGger:DELay? [{MIN|MAX|DEF}]
Sets the delay between the trigger signal and the first measurement.
Parameter
Typical Return
0 to ~10
00 seconds (~1 μs steps), default
1 s
+1.00000000E-06
Return five DC voltage measurements. There is a 2 second delay before each
measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL 2
READ?
Typical Response
:
+3.07761360E-03,-1.16041169E-03,+5.60585356E-06,+1.21460160E-04,+2.85898
531E-04
Due to internal quantization, the actual delay that you set may be slightly different
than your specified value. The
increment is approximately 1 μs. Use the query to
determine the exact delay. For example, if you send TRIG:DEL 500 ms and then
TRIG:DEL? the actual delay may be +5.00000753E-01.
By default, TRIGger:DELay:AUTO is ON. The instrument automatically determines
the delay based on function, range and integration time (see Automatic Trigger
Delays). However, you may need to set a delay longer than the automatic delay for
long cables, high capacitance or high impedance signals.
If you specify a trigger delay with this command, that delay is used for
all
functions
(
except
CONTinuity and DIODe) and ranges. The CONTinuity and DIODe tests
ignore the trigger delay setting.
If you have configured the instrument for more than one measurement per trigger
(SAMPle:COUNt >1), the delay is inserted after the trigger and between consecutive
measurements.
The instrument selects automatic trigger delay after a Factory Reset.