![SIGLENT TECHNOLOGIES CO.,LTD SDM3045X Remote Manual Download Page 82](http://html1.mh-extra.com/html/siglent-technologies-co-ltd/sdm3045x/sdm3045x_remote-manual_1267548082.webp)
Remote Manual
SIGLENT
SDM3045X 80
2.9.30 [SENSe:]VOLTage:{AC|DC}:RANGe:AUTO {OFF|ON|ONCE}
[SENSe:]VOLTage:{AC|DC}:RANGe:AUTO?
Disables or enables autoranging for AC and DC voltage measurements and for DC ratio
measurements.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for AC and DC measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0(OFF)or1(ON)
Configure DC voltage measurements and perform an immediate autorange.
Make and read two measurements:CONF:VOLT:AC
VOLT:AC:RANG:AUTO ONCE
SAMP:COUN 2
READ?
Typical Response
:
+8.36187601E-03,+8.34387541E-03
Under the condition of opening in automatic adjustment range, the instrument is
based on the input signal selection range.
This parameter is set to its default value after a Factory Reset.