SIGLENT Remote Manual
63 SDM3045X
2.9.11 [SENSe:]{FREQuency|PERiod}:VOLTage:RANGe
{<range>|MIN|MAX|DEF}
[SENSe:]{FREQuency|PERiod}:VOLTage:RANGe? [{MIN|MAX|DEF}]
Manually choose frequency/period measurement mode voltage range:
Note
:
This parameter is shared between frequency and period measurements.
Parameter
Typical Return
<range>
:
{600 mV|6 V|60 V|600 V|750V}
,
default 60V
+6.0001
Configures frequency measurements using the 60 VAC range. Make and read two measurements
CONF:FREQ
FREQ:VOLT:RANG 60
SAMP:COUN 2
READ?
Selecting a fixed range ([SENSe:]<
function
>:RANGe) disables autoranging.
If the input voltage is too large for the selected voltage range (manual ranging), the
instrument displays the word Overload on the front panel and returns "9.9E37" from
the remote interface.
This parameter is set to its default value after a Factory Reset or CONFigure
function..