SIGLENT Remote Manual
39 SDM3045X
2.6.4 CONFigure:DIODe
Sets all measurement parameters and trigger parameters to their default values for diode
tests.
Parameter
Typical Return
(
none
)
(
none
)
Configure, make, and read a default diode
measurement:
CONF:DIOD
READ?
Typical Response
:
+1.32130000E-01
The range and resolution are
fixed
for diode tests: the range is 2 VDC.
The FETCh?, READ?, and MEASure:DIODe? queries return the measured voltage,
regardless of its value.
Use READ? or INITiate to start the measurement.
2.6.5 CONFigure:{FREQuency|PERiod}
Sets all measurement parameters and trigger parameters to their default values in the
frequency/period mode.
Parameter
Typical Return
(none)
(
none
)
Configure, make, and read a default
frequency measurement
:
CONF:FREQ
READ?
Typical Response
:
+7.7901
If the input signal is greater than can be measured on the specified manual range, the
instrument
displays
the
word
Overload
on
front
panel
and
returns
"+9.9037" from the remote interface.