2.8.6 MEASure:TEMPerature?[{RTD|THER|DEFault}[,{<type>|DEFault}]]
Sets all measurement parameters and trigger parameters to their default values and
immediately triggers a measurement. The results are sent directly to the instrument's
output buffer.
Parameter
Typical Return
<probe_type>: {RTD|THER}
,
default THER. The
default command can only choose the built-in
sensor manufacturers, does not support user
defined sensor selection.
(
none
)
<type>: PT100(only possible value for RTD ) or
{BITS90|EITS90|JITS90|KITS90|NITS90|RITS9
0|SITS90|TITS90}(The sensor can be selected
for THER)
Configure the RTD measurement. Then read out the measuring results:
MEAS:TEMP? RTD,PT100
Typical Response
:
-2.0002
To change temperature units, use UNIT:TEMPerature.
If the input signal is greater than can be measured, the instrument displays the word
Overload on front panel and returns "9.9E37" from the remote interface.