Remote Manual
SIGLENT
SDM3045X 76
[SENSe:]VOLTage Subsystem
This subsystem configures AC voltage measurements and DC voltage measurements.
Command Summary
[SENSe:]VOLTage:{AC|DC}:NULL[:STATe]
[SENSe:]VOLTage:{AC|DC}:NULL:VALue
[SENSe:]VOLTage:{AC|DC}:NULL:VALue:AUTO
[SENSe:]VOLTage:{AC|DC}:RANGe
[SENSe:]VOLTage:{AC|DC}:RANGe:AUTO
[SENSe:]VOLTage[:DC]:NPLC
[SENSe:]VOLTage[:DC]:IMPedance
2.9.26 [SENSe:]VOLTage:{AC|DC}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:] VOLTage:{AC|DC}:NULL[:STATe]?
Enables or disables the null function for AC or DC voltage measurements.
Note:This parameter is not shared between AC and DC measurements. The parameters
are independent for ACand DC measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default OFF
0(OFF)or1(ON)
Configure AC voltage measurements, using the null function to subtract 100
mV from the measurements. Make and read two measurements:
CONF:VOLT:AC
VOLT:AC:NULL:STAT ON;VAL 100mV
SAMP:COUN 2
READ?
Typical Response
:
+1.0300,+1.0300
Enabling the scaling function also enables automatic null value selection
([SENSe:]VOLTage: {AC|DC}:NULL:VALue:AUTO ON).
To set a fixed null value, use this command:
[SENSe:]VOLTage:{AC|DC}:NULL:VALue.
The instrument disables the null function after a Factory Reset or CONFigure
function.