91 SDM3045X
2.11 TRIGger Subsystem
The TRIGger subsystem configures the triggering that controls measurement acquisition.
Command Summary
TRIGger:COUNt
TRIGger:DELay
TRIGger:DELay:AUTO
TRIGger:SLOPe
TRIGger:SOURce
2.11.1 TRIGger:COUNt {<count>|MIN|MAX|DEF|INFinity}
TRIGger:COUNt? [{MIN|MAX|DEF}]
Selects the number of triggers that will be accepted by the instrument before returning to
the "idle" trigger state.
Parameter
Typical Return
1 to 1,000,000 or continuous (INFinity).
Default 1
+1.0000
For a continuous trigger (INFinity),
the query returns
"9.9E37".
Return ten sets of five DC voltage measurements,
CONF:VOLT:DC
SAMP:COUN 5
TRIG:COUN 10
READ?
Typical Response: -1.85425399E-04
, … (50 measurements)
You can use the specified trigger count in conjunction with a sample count
(SAMPle:COUNt), which sets the number of samples to be made per trigger. In this
case, the number of measurements returned will be the sample count multiplied by
the trigger count.
You can store up to 10,000 measurements in the reading memory of the SDM3055. If
reading memory overflows, new measurements will overwrite the oldest
measurements stored; the most recent measurements are always preserved. No
error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Questionable
Data Register's condition register (see Status System Introduction).
This parameter is set to its default value after a Factory Reset.